Forma Scientific, Inc._______________________________________________Maintenance
6-4
Figure 6-2
Vacuum Relief and Probe Cover Assemblies
6.7 Replacing the Performance Monitor/Alarm Battery
For a consistent and dependable charge, replace the battery
every 2-3 years. Replacement batteries must be of the re-
chargeable type and are available from Forma Scientific, Inc.
Refer to the parts list for stock number and description of the
replacement batteries.
Dispose of the used batteries in a safe manner and in accord
with good environmental practices.
Turn off electrical power to the freezer.
Remove the four screws and lock washers securing the front panel.
Unplug the battery connection and remove the screws securing the batter to the base.
Align the replacement battery pack with the red and black wires in the harness.
Plug in the battery (red to positive and black to negative) and replace the front panel.