Conox
®
Instructions for Use
1370
Manufacturer
Quantium Medical S.L.U.
Av. Ernest Lluch 32, Tower 2,
Tecnocampus, 3rd Floor, Office
3.17, 08302 Mataró, Spain
+34 93 702 19 50
www.quantiummedical.com
Distributor
Fresenius Kabi AG
61346 Bad Homburg, Germany
+4
9
(0)
6
1
7
2
/
686-0
www.fresenius-kabi.com

Table of Contents

Questions and Answers

  • S
    Scott HayesAug 17, 2025
    What to do if Fresenius Kabi Conox Medical Equipment shows 'Lead off'?
    • J
      Joseph MarshallAug 17, 2025
      If the Fresenius Kabi Medical Equipment displays a 'Lead off' error, it could be due to several reasons: * The Patient Sensor may be disconnected: Connect the Patient Sensor. * There might be a poor or contaminated connection: Clean the connection between the Patient Sensor and Patient Cable. * The Patient Cable may be disconnected: Connect the Patient Cable. * A Patient Cable or Sensor could be defective: Replace the Patient Cable or Sensor. If none of these steps resolve the issue, the device itself might be defective, and you should contact your Distributor.
  • P
    Patricia Campbell MDAug 19, 2025
    How to fix 'Impedance too high (TH)' error on Fresenius Kabi Conox Medical Equipment?
    • T
      Tyrone PalmerAug 20, 2025
      If the Fresenius Kabi Medical Equipment indicates that the impedance is too high (TH), it could be due to the following reasons: * The Patient Sensor is not making full contact with the patient's skin: Press the Patient Sensor firmly onto the skin, but avoid pressing the center of the electrode. * The Patient Sensor is defective: Replace the Patient Sensor. If the problem persists, the device itself may be defective, and you should contact your Distributor.
  • S
    Scott CarneyAug 22, 2025
    What to do for excessive artifacts detected on Fresenius Kabi Conox Medical Equipment when SQI is less than 50?
    • M
      Megan ScottAug 23, 2025
      If the Fresenius Kabi Medical Equipment detects excessive artifacts and the SQI is less than 50, consider the following: * Artifacts may be caused by patient eye blinking or movement: Try to minimize patient blinking or movement. * The issue may occur during electrocautery: Note that the device is not designed to work under these conditions. * Electrical activity might be interfering with the EEG, as indicated by the EMG index: The device may be affected by EMC from other devices, such as warming blankets. * The Patient Sensor could be defective: Replace the Patient Sensor. If the problem persists, the device itself may be defective and you should contact your Distributor.
  • R
    Rebekah MolinaAug 25, 2025
    Why is data unavailable on my Fresenius Kabi Conox Medical Equipment due to poor signal quality?
    • K
      kenneth55Aug 26, 2025
      If data is unavailable on your Fresenius Kabi Medical Equipment due to poor signal quality (SQI too low to accurately calculate a qCON index), the following may be the cause: * Artifact caused by patient eye blinking or movement: Try to minimize patient blinking or movement. * Defective Patient Sensor: Replace Patient Sensor. * Defective Patient Cable: Replace Patient Cable. If the problem persists, the device itself may be defective, and you should contact your Distributor.
  • S
    samuelmoyerAug 27, 2025
    What to do when Fresenius Kabi Conox shows 'Battery power low'?
    • C
      Courtney DickersonAug 27, 2025
      If the Fresenius Kabi Medical Equipment displays a 'Battery power low' message, it means the device is operating on backup battery power with only a few minutes of operation remaining. To resolve this, connect an external power supply.
  • E
    eellisAug 30, 2025
    Why is my Fresenius Kabi Medical Equipment operating on backup battery?
    • L
      lmaldonadoAug 30, 2025
      If your Fresenius Kabi Medical Equipment is operating on backup battery unintentionally: * Connect an external power supply. * Replace the external power supply cable. * Replace the external power supply. If the issue continues, the device may be defective; contact your Distributor.
  • L
    Larry AlvarezSep 1, 2025
    Why does the display go black on my Fresenius Kabi Conox?
    • C
      Christopher HerreraSep 1, 2025
      If the display on your Fresenius Kabi Medical Equipment goes black, it could be due to these reasons: * The backup battery is totally discharged: Connect an external power supply. If the issue persists, the device itself may be defective; contact your Distributor.
  • S
    Sue WellsSep 4, 2025
    What does 'Corrupted file' mean on Fresenius Kabi Conox?
    • B
      Barbara WillisSep 4, 2025
      If the Fresenius Kabi Medical Equipment indicates a corrupted file, it means the file cannot be retrieved. Delete the file. If this happens repeatedly, contact your Distributor.
  • D
    David AlvarezSep 6, 2025
    Why haven't the date and time changed on my Fresenius Kabi Conox Medical Equipment?
    • E
      ernest30Sep 6, 2025
      If the date and time have not changed on your Fresenius Kabi Medical Equipment, the device may be defective. Contact your Distributor.
  • M
    Morgan WrightSep 9, 2025
    What does 'Display Micro SD not mounted' mean on Fresenius Kabi Conox Medical Equipment?
    • C
      Casey BaileySep 9, 2025
      If the Fresenius Kabi Medical Equipment displays 'Micro SD not mounted', the internal micro SD of the display has become corrupted. Contact your Distributor.