6 - 32
4) The value is highlighted, indicating that
WKHYDOXHFDQEHFKDQJHG&KDQJHWKH
value by pressing the
or the key,
and then move the cursor to the right by
pressing the
key.
5) After the value is changed, press the
ENT
key.
End of Hold Setting
ESC
Parameter Setting screen
The setting is represents the percentage with
each Ch (component) range regarded as 100%
for both ranges.
When 0 to 1000 ppm is selected as the range,
for example, if 10% FS is selected as hold
setting, the output equivalent to 100 ppm is
output and held irrespective of the measure-
ment value at that time.
Meaning of setting
 ,QVWDQWDQHRXVYDOXHGLVSOD\RIWKHPHDVXUHPHQWFDQQRWEHKHOG2XWSXWRQO\FDQEHKHOG
 ,IVHWYDOXHLVVHOHFWHGIRUKROGLQVWDQWDQHRXV2
2FRQYHUVLRQYDOXHLVFDOFXODWHGDQGKHOG
EDVHGRQWKHVHWYDOXH
 5DQJHLGHQWLILFDWLRQFRQWDFWRXWSXWFDQQRWEHVZLWFKHGHYHQLIWKHUDQJHLVVZLWFKHGGXULQJ
WKHKROG
Description of setting
6) Press
the
ESC
key to return to the param-
eter setting screen.
ENT
ENT
Parameter
Set
Hold
value
Hold
0
to
100%FS
0 0
0
Chl
NOx
0010
%FS
Ch2
S02
020
%FS
Ch3
CO2
015
%FS
Ch4
co
012
%FS
0
Ch5
02
022
%FS
0