6 - 33
Reset input
So long as close, resetting lasts.
At the edge of changing from closing to opening, the average action restarts.
Close (hold at least 1.5 sec.)
Open
Average value reset
This mode is used to clear O
2
average values and O
2
conversion average values and restarts aver-
aging. All average values are reset at a time. The indication value and output value is 0 ppm,
YRORUVRDWWKHWLPHRIWKHUHVHWLQSXW5HIHUWRWKHDYHUDJHSHULSG
Average period
It allows you to set an average period of the
average value of O
2
conversion
and O
2
average.
,WHQDEOHV\RXWRVHWDQDYHUDJHWLPHRIWR
minutes (1-minute step) or 1 to 4 hours (1-hour
step).
&KDQJLQJWKHVHWWLQJUHVHWVWKHDYHUDJHYDOXH
of O
2
conversion and O
2
average value. (Press-
ing the
ENT
validates the resetting only for
components whose setting was changed.)
Response time
The response time of the electrical system can
be changed.
Setting is available by components.
Note) It does not provide exact seconds for
the setting time, but it gives a guide
of the setting time.
7KHVHWWLQJYDOXHFDQEHPRGL¿HGDV
requested by the customer.
~----~-------------------------
Parameter
Select
Ch
No.
Response
Time
~
Chl
NOx
10
sec
Ch2
SO2
20
sec
Ch3
CO2
15
sec.
Ch4
co
12
sec.
Ch5
02
22
sec.
Parameter
Select
Ch
No
Average
Period
~
Ch9
~NOx
01
hour
Chl0
~SO2
01
hour
Chll
36CO2
01
hour
0
Ch12
IW.
02
01
hour