COMPLETION OF SENSOR CALIBRATION
K0553 Revision D 117 [EN] English
Completion of sensor calibration
After completion of all calibration adjustment procedures, carry out the following:
1) Make sure the calibration standard and the ADTS are at atmospheric pressure. Disconnect
the calibration standard from the ADTS.
2) If there is no further calibration or testing required, exit menus, select standby or switch off
the ADTS.