SOFTWARE UPDATES
K0553 Revision D 130 [EN] English
3) Switch-on the ADTS Touch (1).
4) Insert the USB memory stick into the ADTS Touch USB port (2).NOTE: The USB stick must not
be removed once the install process has started.
Figure 5-3: Software updates-ADTS Touch
5) On the “Dashboard”, touch “Tools”. The tools menu opens.
6) Touch “Calibration”. A numbered keypad is displayed.
7) Enter your software update PIN and touch the “tick” icon. The “Software Upgrade” menu is
displayed containing two items: ADTS TOUCH and ADTS.
ADTS TOUCH
- APPLICATION. You will be asked to confirm the software upgrade “Yes” or “No”.
- OPERATING SYSTEM. You will be asked to confirm the software upgrade “Yes” or “No”.
ADTS
- MAIN CODE. You will be asked to confirm the software upgrade “Yes” or “No”.
- BOOT CODE. You will be asked to confirm the software upgrade “Yes” or “No”.
8) Once you have selected “Yes”, follow the on-screen instructions.
9) Selecting “No” closes the dialog without making any changes.
Downloading ADTS manuals
1) Go to www.gemeasurement.com. The “GE Measurement & Control” page is displayed.
2) On the top level menu bar, click Support. Menu items under that header are shown.
3) Click Download Center. The “Download Center” page is displayed showing the search
panels.

Table of Contents

Other manuals for GE ADTS542F

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE ADTS542F and is the answer not in the manual?

GE ADTS542F Specifications

General IconGeneral
BrandGE
ModelADTS542F
CategoryTest Equipment
LanguageEnglish

Related product manuals