Why does my GE Druck DPI612 have an excessive leak rate (pressure or vacuum)?
- KkmartinAug 17, 2025
An excessive leak rate in your GE Test Equipment could be due to several reasons: * The system might be incorrectly connected (DPI612 to DUT). Ensure that connecting accessories are fitted according to instructions and that the correct sealing method is in use. * The DUT (Device Under Test) may have a leak. Isolate the DUT from the DPI612 and check for leaks in both independently. * There might be a worn pressure seal on the outlet port ('O' ring & backup ring). Inspect the seal on the pressure port for signs of wear and replace if necessary. * The Release Valve might not be fully closed. Make sure the Release Valve is firmly closed.











