Page 14 of 105 [EN] English – 109M4017 Revision -
Wipe the battery terminals with a clean dry cloth if they become
dirty.
If you find rust, a bad odour, overheating and/or other irregularities
when using the battery pack for the first time, return it to your
supplier or vendor.
For further information contact the nearest distributor or
representative.
Retain the original product literature for future reference.
PRESSURE WARNINGS 1.7
It is dangerous to attach an external source of pressure to a
DPI612 pressure station. Use only the internal mechanisms to set
and control pressure in the pressure station.
To prevent a dangerous release of pressure, isolate and bleed the
system before disconnecting a pressure connection.
To prevent a dangerous release of pressure, make sure that all the
related pipes, hoses and equipment have the correct pressure
rating, are safe to use and are correctly attached.
To prevent damage to the DPI612 calibrator, only use it within the
specified pressure limits.
Do not exceed the maximum pressures stated in the appropriate
component manual for the unit under test.
Reduce pressure at a controlled rate when venting to atmosphere.
Carefully de-pressurize all pipes to atmospheric pressure before
disconnecting and connecting to the unit under test.
Always wear appropriate eye protection when working with
pressure.