User Manual Druck DPI612
Page 52 of 105 [EN] English – 109M4017 Revision -
Note: To discard the hydraulic fluid, obey all the local health and
safety procedures.
2.3.8 Drain all the Hydraulic Fluid
In some conditions, it is necessary to fully drain the hydraulic fluid
from your DPI 612 hFlexPro pressure calibrator; for example:
if you are using water and the storage or operating temperature is
going to be less than 4°C (39°F)
if there is a long period of storage
if there is unwanted material in the hydraulic fluid
Preparation
To drain the instrument, we recommend these items:
the applicable skin and eye protection
a container that is large enough to hold the hydraulic fluid and
prevent contamination of the work surface
applicable materials to make sure the instrument and the area
stay clean;