User Manual Druck DPI612
Page 85 of 105 [EN] English – 109M4017 Revision -
A Druck DPI612 calibrator device driver available as a
download from the website www.gemeasurement.com
.
4.4.1 Sequence to Upload and Download file
Table 4-1
Step Procedure
1
Connect the USB cable (See
Section 1.10,
Accessories)
to the Druck DPI612 calibrator.
2
Connect the lead to the USB port on the computer
that has the calibrator manager installed.
3 Use 4 Sight to set-up the procedure and create a
work order for the device.
The procedure includes the parameters for the
calibration, the number of test points, the relationship
and pass/fail tolerance.
4
Use the Download button in calibrator manager to
Download the file to the Druck DPI612 calibrator. A
communications symbol will be displayed at the
bottom of the screen.
5
Select:
DASHBOARD >> DOCUMENTING >> RUN PROCEDURE
6 In the Results window, select the filename specified in
calibrator manager.
7 Enter the User ID and the DUT Serial Number.
8
Tap on the Start button . The procedure sets up the
necessary Channel options e.g. mA and Volts.
9
Use the Take Reading button at each point specified
by the procedure. A prompt show for each point.
10 When all the readings are complete, tap on the Exit
Button . Look at the results on the display
(As found/As Left).
11
To complete the process, use the calibrator manager
to Upload the file back into the 4 Sight database.