EasyManua.ls Logo

GE Druck DPI612 - Dirt Moisture Trap 100 Bar (P;N Io620-Idt622); Specification; Pressure Connections; Operation

GE Druck DPI612
105 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
User Manual Druck DPI612
Page 96 of 105 [EN] English – 109M4017 Revision -
Dirt Moisture Trap 100 bar (P/N IO620-IDT622) 6.2
6.2.1 Specification:
Maximum working pressure: 100bar / 1500psi pneumatic.
Important Notes:
This device is designed for use on the DPI612 PFlexPro100 bar
pneumatic pressure calibrator.
It should be removed from the calibrator during transit.
It should be inspected internally for dirt or moisture before and after
use.
Orientation: Use only in the vertical position.
6.2.2 Pressure connections:
Lower: To DPI 612 pFlexPro test port.
Upper: AMC quick-connect (for use with all DPI612 adaptor sets)
Materials of Construction: 316L and 303 Stainless steels, 6082
Aluminium, Nitrile seals, P.T.F.E.
6.2.3 Operation:
1. Check for moisture or dirt particles within the aluminium
chamber before and after use.
If found, dismantle and clean as instruction below.
2. Fit the Dirt Trap to the Test Station of the DPI 612 pFlexPro
pressure calibrator.
Hand-tight only no spanners / wrenches are required.
3. Fit AMC adaptor to suit item under test. ( hand-tight only)

Table of Contents

Related product manuals