Table 5-3 Current (source) error limits
Applied
mA
Calibrator
uncertainty
(mA)
Permitted
DPI612 error
(mA)
0.2
0.00008
0.0013
6 0.00023 0.0017
12
0.00044
0.0020
18
0.0065
0.0024
24
0.0012
0.0028
PROCEDURES: DC mV/Volts (measure) 5.7
Do the procedure as follows:
1. Connect the applicable calibration equipment (See Table 5-1).
2. Let the equipment get to a stable temperature (minimum: 5
minutes since the last power on).
3. Use the calibration menu (See Section 5.4) to do a three-point
calibration.
(-FS, Zero and +FS) for the applicable set of ranges:
mV (measure) ranges Volts (measure) ranges
200 mV
20 V
2000 mV
30 V
4. Check the calibration is correct:
Select the applicable Millivolts or Voltage (measure) function.
Apply the input values that are applicable to the Calibration:
mV: -2000, -1000, -200, -100, 0 (short circuit)
mV: 0, 100, 200, 1000, 2000
Volts (V): -30, -21, -20, -10, -5, 0 (short circuit)
Volts (V): 0, 5, 10, 20, 21, 30.
Check that the error is within limits (See Table 5-4 and Table
5-5).