User Manual Druck DPI612
Page 91 of 105 [EN] English – 109M4017 Revision -
Table 5-4 Millivolts (measure) error limits
Applied
mV
uncertainty
(mV)
DPI612 error
(mV)
±1000 0.040 0.0940
±100 0.0040 0.0110
0 (short circuit) 0.0036 0.0070
Table 5-5 Voltage (measure) error limits
Applied
V
Calibrator
uncertainty
(V)
Permitted
DPI612 error
(V)
±20 0.00031 0.00118
±10 0.00016 0.00089
±5 0.00008 0.00075
0 0.000024 0.00060
PROCEDURES: DC Volts (source) 5.8
Do the procedure as follows:
1. Connect the applicable calibration equipment (See Table 5-1).
2. Let the equipment get to a stable temperature (minimum: 5
minutes since the last power on).
3. Use the calibration menu (See Section 5.4) to do a two-point
calibration for the applicable range:
Volts (source) range
10 V
24 V