User Manual Druck DPI612
Page 23 of 105 [EN] English – 109M4017 Revision -
DPI 612 pFlexPro (100 bar) Pump
When you have set the operation to pressure or vacuum (
See
Section
1.9.4), seal the system (See Section 1.9.3) and use the
pump to set the necessary pressure or vacuum.
You can then make the last adjustments with the volume
adjuster (See Section 1.9.6).
100 bar Pump
DPI 612 hFlexPro (1000 bar) Pump
When you have set the operation to pressure or vacuum
(See Section 1.9.4), seal the system (See Section 1.9.3) and
use the pump to set the necessary pressure or vacuum.
You can then make the last adjustments with the volume
adjuster (See Section 1.9.6).
1000 bar
Pump
1.10.6 Volume Adjuster
DPI 612 pFlex (20 bar) Volume Adjuster
This control increases or decreases the pressure/vacuum.
Before you seal the system (See Section 1.9.3), turn this
control to the necessary position:
• For equal adjustment, turn it to the middle of
its range.
• For maximum adjustment, turn it fully
clockwise or counter clockwise.
When you have set the necessary pressure or vacuum
with the pump (See Section 1.9.5), use the volume adjuster
to make the fine adjustments.
20 bar
Volume
Adjuster