_________________________
Page 26 of 33 © 2007 General Electric Company 2023491-201 Rev B
WEIGHT
Model Weight with Batteries and Pads
9300 3.20 kg (7.0 lb)
ENVIRONMENTAL OPERATION AND STANDBY CONDITIONS
Atmosphere Condition
Temperature 0°C to 50°C (32°F to 122°F)
Humidity 5% to 95% (non-condensing)
Pressure 57kPa (4,572m / +15,000ft) to 103kPa (-152m / -500ft)
SHIPMENT AND TRANSPORT ENVIRONMENTAL CONDITIONS (for up to 1 week)
Atmosphere Condition
Temperature -30°C to 65°C (-22°F to 149°F)
Humidity 5% to 95% (non-condensing)
Pressure 57kPa (4,572m / +15,000ft) to 103kPa (-152m / -500ft)
PADS
• Self-adhesive, disposable defibrillation pads
• Minimum combined surface area: 228cm2
• Extended length of lead wire: 1.3m
LITHIUM BATTERY SPECIFICATIONS
• Output voltage: 12VDC (max)
• Batteries are non-rechargeable
• Lithium contents: 9.2g (max)
• Check local regulations for disposal information
Model Estimated Shelf Life Typical Shocks
2023681 Lithium 5 Years Up to 290 shocks
The battery operating life depends on the type of battery, actual usage and environmental factors.
RECHARGEABLE BATTERY SPECIFICATIONS
• Battery Voltage: 11.1V
• Chemistry: Lithium-ion. Refer to local regulations.
• Compatibility: Responder AED Pro Model 2023440
• Battery Capacity: 60 shocks minimum (100 shocks typical) or 3 hours minimum (6 hours typical) of ECG display time.
• Battery Charge Time: 3 hours for stated capacity, 4.5 hours to fully charge completely depleted battery.
• Battery Standby: 6 months
• Battery Life: 2.5 years or 300 Battery charge-discharge cycles, whichever comes first.
• Battery Weight: 1 lb. 3 oz