ÄKTA start System Cue Card, 29-0240-42 AC 8
Evaluating a run
Save Result as BMP image
The system generates a BMP le of the result which can be
viewed under any OS that has an image viewer supporting
BMP. The chromatogram is plotted as UV abs (mAU) vs
Volume (ml), along with Fractionation marks.
Note: BMP le is generated only for the last 4 hours of run.
Evaluating a run using UNICORN start
Evaluation module of UNICORN start is used to manage
and evaluate the results from chromatography runs. For
instructions to evaluate results that are generated using
ÄKTA start, refer to the UNICORN start 1.0 User Manual.
Save Result to USB memory stick
Import Result to UNICORN start
The result le stored on the USB stick can be imported to
UNICORN start for analysis, printing and generating a report.
Quick start methods
Quick start contains “ready to run” methods like Anity,
Ion exchange, Gel ltration and Desalting. The run
parameters are pre-set in the method. Details of sample
volume need to be provided before starting the run.
If required, the run parameters can be changed (Edit
run in Run view screen) when the run is in progress.
Use appropriate columns as indicated in the template
names. E.g., use 1 ml HiTrap™ column in case of
selecting the Quick start AC/IEX step 1 ml HiTrap, or 5 ml
column when selecting AC/IEX step 5 ml HiTrap.
Quick start workow
Connect USB stick to
ÄKTA start
Select chromatography
method (for method run)
Manual run / Method run
In the Run parameters or
Manual run screens, tick the
checkbox Save Result to USB
Import and view Result file on
UNICORN start or BMP image
on any other OS.
UNICORN start
File : Import
Import Result from USB stick:
• Locate USB drive
• Locate and select the result
file. Select multiple files using
Shift or Ctrl keys.
Open
Select destination folder
for result
Evaluation
Import
Pre defined
UNICORN start
System control
Method run
Quick start
AC / IEX / DS / GF
Note: The result les will be saved in the GE folder which
is automatically created by the instrument once the USB
memory stick is plugged in. Only 10 results can be stored
in the GE folder. To save further results, transfer the result
les to another folder, PC or rename the GE folder.
Method run
Select the desired
technique
Quick start
Set Sample volume
Tick the checkbox
Save Result to USB
Edit run to alter
run parameters
ÄKTA start
EndRun
Exit

Other manuals for GE akta start

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE akta start and is the answer not in the manual?

GE akta start Specifications

General IconGeneral
BrandGE
Modelakta start
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals