ÄKTA start System Cue Card, 29-0240-42 AC 8
Evaluating a run
Save Result as BMP image
The system generates a BMP le of the result which can be
viewed under any OS that has an image viewer supporting
BMP. The chromatogram is plotted as UV abs (mAU) vs
Volume (ml), along with Fractionation marks.
Note: BMP le is generated only for the last 4 hours of run.
Evaluating a run using UNICORN start
Evaluation module of UNICORN start is used to manage
and evaluate the results from chromatography runs. For
instructions to evaluate results that are generated using
ÄKTA start, refer to the UNICORN start 1.0 User Manual.
Save Result to USB memory stick
Import Result to UNICORN start
The result le stored on the USB stick can be imported to
UNICORN start for analysis, printing and generating a report.
Quick start methods
Quick start contains “ready to run” methods like Anity,
Ion exchange, Gel ltration and Desalting. The run
parameters are pre-set in the method. Details of sample
volume need to be provided before starting the run.
If required, the run parameters can be changed (Edit
run in Run view screen) when the run is in progress.
Use appropriate columns as indicated in the template
names. E.g., use 1 ml HiTrap™ column in case of
selecting the Quick start AC/IEX step 1 ml HiTrap, or 5 ml
column when selecting AC/IEX step 5 ml HiTrap.
Quick start workow
•
•
Note: The result les will be saved in the GE folder which
is automatically created by the instrument once the USB
memory stick is plugged in. Only 10 results can be stored
in the GE folder. To save further results, transfer the result
les to another folder, PC or rename the GE folder.