SpecificationTool
Certified and calibratedBalance
The balance must be able to weigh enough solution
for the maximum flow rate for each system.
6 mm and 3/8”: The balance must be able to
weigh 10 kg/2 minutes collect (maximum flow
rate: 180 l/h)
10 mm and 1/2”: The balance must be able to
weigh 30 kg/2 minutes collect (maximum flow
rate: 600 l/h)
1” SS and PP: The balance must be able to weigh
80 kg/2 minutes collect (maximum flow rate:
2000 l/h)
Three (3) hoses (2 meters each) of the correct in-
ternal diameter
TC connections
Flexible hoses
System documentation is delivered with the system
and is required during installation.
Reference Manuals System
Binder I, II and III
The following chemicals and supplies must be supplied by the user:
SpecificationItem
For example, WFI/MQ purifiedWater of suitable standard for use in the
system
1.0% vol/vol
0.5% vol/vol
0.1% vol/vol
in sealed containers
Acetone solutions
20%Ethanol
pH 4.00 (200 ml)
pH 7.00 (200 ml)
Calibration solutions for pH
Approximately 2, 12 and 110 mS/cmConductivity solutions
ÄKTAprocess Site Preparation Guide 29-1396-55 AA 11
2 Site preparation overview

Other manuals for GE AKTAprocess

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE AKTAprocess and is the answer not in the manual?

GE AKTAprocess Specifications

General IconGeneral
BrandGE
ModelAKTAprocess
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals