2 Site preparation overview
Introduction
Before the product is delivered, you must prepare the site according to the instructions
in this chapter.
Select the site
Select a site that complies with national and international requirements for biosafety.
WARNING
Spread of biological agents. The operator must take all necessary
actions to avoid spreading hazardous biological agents. The facility
must comply with the national code of practice for biosafety.
General site requirements
The following requirements must be fulfilled at the location where the ÄKTAprocess
system will be installed:
•
ÄKTAprocess is intended for indoor use only
•
The system units must not be installed in a dusty environment
•
The system units must not be exposed to sources of heat, such as direct sunlight
•
The system units must not be exposed to strong magnetic or electrical fields
•
The system units must not be exposed to vibrations
•
The system units must not be exposed to corrosive gas
ÄKTAprocess Site Preparation Guide 29-1396-55 AA 5