EasyManuals Logo

GE AKTAprocess Site Preparation Guide

GE AKTAprocess
16 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #10 background imageLoading...
Page #10 background image
Mains and neutral conductor
minimum area
Mains current
up to
16 AWG or 1.5 mm
2
10 A
14 AWG or 2.5 mm
2
16 A
12 AWG or 4 mm
2
25 A
Computer requirements
An external computer is not required. ÄKTAprocess is provided with a built-in industrial
computer.
User-supplied accessories and
consumables
The following equipment must be supplied by the user:
SpecificationTool
Used for unpacking the system13 mm wrench (or ratchet
wrench with 13 mm socket)
Used for unpacking the systemElectrical TX25 screwdriver or
power driver with a hexalobu-
lar TX25 bit
Used for unpacking the systemSafety knife
Used during installation, operation and maintenancePersonal Protective Equip-
ment (PPE)
as required, (e.g., gloves, lab
coats, etc.)
Suitably sized vessels to allow collection of liquid for
10 minutes at maximum flow rate, and a suitably
dimensioned waste system.
Buffer and waste vessels
10 ÄKTAprocess Site Preparation Guide 29-1396-55 AA