GENERAL
ELECTRIC
MEDICAL
SYSTEMS
AMX-li
X-RAY
UNIT
10-377
3-3
MA
and
Filament
Protection
initial
conditions:
Main
power
circuit
breaker-OFF
X-ray
generator
standby
switch—OFF
Charger
cord
disconnected
SM
A0654E
OPERATION
RESULTS
EXPECTED
SERVICE
REFERENCE
la.
Remove
the
jumper
(with
banana
plugs)
from
terminals
MA+
&
MA-
on
panel
3.
install
an
accurate
0
to
150
ma,
DE
millameter
with
plus
lead
in
the
MA+
end
and
the
minus
lead
in
the
MA-.
b.
Connect
a
O
to
150
v,
iron
vane
type,
AC
voltmeter
across
3TB1-18
and
3TB1-12
to
read
XS-XC
Filament
Primary
Voltage.
c.
Turn
the
x-ray
tube
port
up,
close
collimator
blades
and
cover
collimator
port
with
a
1/8”
(3.175
mm)
thick
lead
sheet.
d.
Turn
circuit
breaker
and
standby
switch
ON.
Select
80
KVP,
200
MAS,
(2-second
exposure).
소소
e.
Take
an
exposure
and
read
the
ma
meter.
Do
not
make
more
than
one
expo-
sure
per
min.
속어.
Make
another
exposure
and
read
the
voltage
across
3TB1-18
and
3TB1-12.
At
the
end
of
the
exposure.
g.
Press
the
Rotor
button
only
and
read
voltage
across
3781-18
and
3TB1-12.
MA
should
be
between
95
and 105
MA.
AC
Filament
Primary
Voltage
on
3781-18
and
37B1-12
should
not
change
more
than
1.0-v.
RMS
between
steps
f
and
g.
If
MA
is
not
100,
adjust
3A2R34
to
obtain
100
MA
X-ray
tube
current.
Refer
to
Chapter
4,
Sec.
1-3
or
Chapter
4,
Part
Il,
Sec.
Il.
Adjust
3A2R30
to
obtain
the
same
reading
as
in
step
f.
Refer
to
Chapter
4,
Sec.
1-3
or
4-11-28.
5-9