AMX-II
X-RAY
UNIT
1-981
OM
A0654E
Technic
Factor
Deviations
The
maximum
deviation
from
the
selected
setting
during
radiographic
exposure
is
as
follows:
SELECTION
AND
ACCURACY
AS
STATED
RANGE
IN
(MEDIAN*)
+
VARIATION**
MAS
1
MAS
(+/-
40%
+
.35MAS)
+/-
.27
MAS
1.5
MAS
(+/-
33%
+
.30MAS)
+/-
.27
MAS
2
MAS
|
(+/-
21%
+
.32
MAS)
+/-
.28
MAS
2.5
MAS
to
4
MAS
(+/
14%
+
.39
MAS)
+/-
.30
MAS
5
MAS
to
8
MAS
(+/-
8%
+
.40
MAS)
+/-
.32
MAS
10
MAS
(+/-
7%
+
.40
MAS)
+/-
.32
MAS
12
MAS
to
300
MAS
-7%
to
+10%
*
Represents
the
range
of
possible
median
values
for
any
given
machine.
**Represents
the
range
of
possible
variation
between
exposures
at
fixed
technic
for
any
given
machine.
KVP
50
KVP
to
110
KVP
(+
(10
KVP
+7%))
(-(5
KVP
+
7%))
of
dial
reading
after
first
20
milliseconds
of
exposure
50
KVP
to
95
KVP
(+
(20
KVP
+7%))
'(-(5
KVP
+7%))
of
dial
reading
during
first
20
milliseconds
of
exposure.
100
KVP
to
110
KVP
(+
(25
KVP
+7%))
(-(5
KVP
+
7%))
of
dial
reading
during
first
20
milliseconds
of
exposure.
Conditions
required
to
achieve
the
above
accuracies:
1.
The
Battery
Condition
Meter
must
be
in
either
of
the
two
green
bands.
2.
Allow
a
minimum
interval
of
15
minutes
after
charge
completion
with
key
switch
in
“ON”
or
“CHARGE”
position,
and
main
power
circuit
breaker
“ON”,
before
making
an
exposure.
14