GENERAL
ELECTRIC
MEDICAL
SYSTEMS
AMX-It
X-RAY
UNIT
7-1576
'
SM
A0654E
8.
Turn
the
charger
off
when
battery
voltage
reaches
110
volts.
Turn
circuit
breaker
OFF,
remove
the
lamp
load
in
series
with
DC
ammeter
and
reconnect
ammeter.
Be
sure
meter
polarity
is
correct.
9.
During
the
remainder
of
the
charge
it
will
be
necessary
to
monitor
and
record
the
charge
current
at
1/2
hour
intervals
so
as
to
calculate
the
total
ampere-hour
charge
input
to
the
battery.
Prepare
a
table
as
shown
below
to
record
the
data.
The
table
shows
typical
data
during
the
initial
charge
of
a
new
battery.
CHANGE
INPUT
TO
BATTERY
DATE
&
HOUR
INTERVAL
BATTERY
CHARGE
CURRENT
INTERVAL
ACCRUED
TIME
FROM
STEP
10
VOLTAGE
END
OF
INTERVAL
AMP-HOURS
AMP-HOURS
_
6/27
12:07
0.0
117.4
_
2.5
12:37
0.5
122.1
2.1
1.05
1.05
13:07
1.0
122.6
2.5
1.25
2.30
13:37
1.5
123.0
2.4
1.2
3.50
14.07
2.0
123.4
2.5
1.25
4.75
14:37
2.5
123.8
2.4
1.2
5.95
15.07
3.0
124.2
2.4
1.2
7.15
15:37
3.5
124.5
2.4
1.2
8.35
16:07
4.0
124.8
2.3
1.15
9.50
16:37
4.5
125.1
2.3
1.15
10.65
17:07
5.0
125.3
2.3
1.15
11.80
Overnight
Shutdown
6/28
98:00
5.0
119.4
.
2.7
——
08:30
5.5
125.3
2.3
1.15
12.95
09:00
6.0
125.7
2.2
1.1
14.05
09:30
6.5
126.1
2.3
1.15
15.20
10:00
7.0
126.3
2.2
1.1
16.30
10:30
7.5
127.1
2.2
1.1
17.40
11:00
8.0
127.6
2.1
1.05
18.45
C/10
Charge
Rate
11:15
8.0
126.2
1.3
0.65
19.10
11:45
8.5
129.1
1.3
0.65
19.75
12:00
8.75
131.5
1.3
0.325
20.075
‘Charger
Cut-off;
Overcharge
12:15
8.75
130.5
1.3
12:30
9.0
133.5
1.3
0.325
20.40
13:00
9.5
139.8
1.3
0.65
21.05
3-18