Starting a case using default settings
Case Defaults contain ve case type selections. Each case type has preset values for Ideal Weight, Age, and
Volume Apnea Alarm. The rst four default case types are congured and named by the Super User. The fth
default case is Last Case.
1. Set the Bag/Vent switch to Bag.
2. Select Start Case.
The Case Defaults selection shows the rst preset case type
Ideal Weight, Age, CO2 Alarms, and Volume Apnea Alarm show the default settings that correspond to
the case type shown
3. Verify or change the Case Defaults selected.
4. Verify the settings are clinically appropriate.
5. Select Start Case Now. Gas ow starts.