SIMV PCV-VG mode
• Delivers a set rate of pressure controlled breaths with a guaranteed volume
• The ventilator will adjust the inspiratory pressure needed to deliver the set tidal volume breath-by-breath so that
the lowest pressure is used.
• The patient’s compliance is determined from the volume controlled ventilation breath and the inspiratory
pressure level is then established for the next PCV-VG breath.
SIMV PCV-VG mode settings:
• TV
• RR
• Flow Trigger
• Psupport
• PEEP
• Tinsp
• Pmax
• Trig Window
• End of Breath
• Rise Rate
• PSV Rise Rate
Tinsp
Texp
Trigger
window
Trigger
window
Flow
TV TV
Time
Pressure
Time
Variable pressure to
deliver desired TV
Tinsp Texp
Peep
33