35
08 Special features
ecoFLOW overview
The ecoFLOW option provides a split screen view that
shows the approximate minimum O
2
ow to maintain a
preset FiO
2
value. It also shows the approximate agent
used per hour and the cost.
The split screen shows the Paw gauge in the upper area
and the ecoFLOW gauge in the lower portion of the
screen. The ecoFLOW gauge consists of a fresh gas ow
tube, an agent ow indicator, and related parameters.
Below the fresh gas ow tube is the measured total O
2
ow to the patient and the calculated FiO
2
ow value. The
FiO
2
ow value is based on the FiO
2
setting specied by a
Super User. This is the minimum O2 ow needed to deliver
a preset inspired O
2
concentration. O2 Total ow greater
than the Fi25 O
2
ow represents excess ow which can be
reduced, while still maintaining desired FiO
2
preset value.
The FiO2 ow value is specic to each patient and case.
The agent ow indicator shows the amount of liquid agent
ow as related to the fresh gas setting. The calculated
cost of the agent shows above this indicator. This cost is
based on agent ow and the values entered in the Super
User mode.
I/min ml/h
02 Total
Sev
€ 6.00 /h
Fi 35
Fi 35 02 flow
0.50
1
2
200
9
20
15
3
Total Flow
0.55
8
Agent