PSVPro™ Ventilation Mode
•  Pressure supported ventilation with apnea backup
•  Clinician sets the Psupport and PEEP levels, the patient establishes the rate, inspiratory ow and inspiratory time
•  Tidal volume is determined by the pressure, lung characteristics and patient eort
•  Apnea backup mode (SIMV PCV) is provided if the patient stops breathing
•  In backup mode, the Backup Mode active alarm is shown until PSVPro™ Ventilation Mode is reinstated or another 
ventilation mode is selected
PSVPro mode settings:
•  Psupport
•  PEEP
•  Trig Window
•  Flow Trigger
•  End of Breath
•  Pmax
•  Backup Time
•  Pinsp
•  RR
•  Tinsp
•  Rise Rate
•  Exit Backup
Time
Pressure
Peep
CPAP + PSV mode
•  Provides a constant support pressure for spontaneously breathing patients once the ventilator senses an 
inspiratory eort
•  The patient establishes the rate, inspiratory ow, and inspiratory time
•  The tidal volume is determined by the pressure, lung characteristics, and patient eort.
•  Spontaneous breaths that occur are indicated by a color change in the waveform
CPAP + PSV settings:
•  Psupport
•  PEEP
•  Flow Trigger
•  End of Breath
•  Pmax
•  Pinsp
•  Minimum RR
•  Tinsp
•  Rise Rate
34
Time
Pressure
Peep
T
insp
T
insp