EasyManua.ls Logo

GE DewPro MMY30 - Chapter 6: Technical Specifications; MMY30 Specifications; Sensing Element; Measurement Range

GE DewPro MMY30
47 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Technical Specifications 6-1
November 2005
MMY30 Specifications
Sensing Element Planar sensor, aluminum oxide capacitance principle
Measurement Range - 90°C to +10°C (-130°F to +50°F) dew point temperature.
0 to 10, 0 to 100, 0 to 1000 ppm
v
(fully adjustable with integral
display)
Recommended
Recalibration Cycle
12 months, depending on the application
Calibration Accuracy ±2°C (±3.6°F) dew point at 25°C (77°F)
Maximum Sensor Relative
Humidity
50% at dewpoint temperatures >0°C (32 °F)
Operating and Storage
Temperature
-40°C to +50°C (-40°F to +122°F)
Air Bleed Off at 7 bar (100
psig)
Approximately 28 sl/h (1 SCFH)
Maximum Operating
Pressure
31 bar, 3.1 MPa (450 psig)
Helium Leak-Rate <10
-6
mbar l/s
Output 4 to 20 mA; 16 µA resolution
Flow Block 316 stainless steel with G ½ thread (DIN ISO 228) or ½” (12.7 mm)
MNPT thread
Wrench Width for Flow
Block
42 mm (1 5/8”)
Electronics Microcontroller operated
Moisture Unit Dew point temperature in °C or °F, ppm
v
Power Supply 24 VDC nominal, 12 to 28 VDC range
Protection Type 4X (IP 67)
Weight 1.5 kg (3.2 lbs)
European Compliance Complies with EMC Directive 89/336/EEC and PED 97/23/EC for
DN<25

Table of Contents

Related product manuals