4 DINAMAP® PRO Series 100-400V2 Monitor Pre-Service & Calibration Manual Revision A
2018554-001
Avoid storing batteries at temperatures above 77° F (25° C). High
storage temperatures can dramatically increase the self-
discharge rate of battery.
1.3.5 Checking Calibration
Perform the test procedures described in 4.6 whenever the accuracy
of any of the parameters is in doubt.
1.4 Clinician Mode
The PRO Monitor provides a special mode named Clinician Mode.
This mode is accessed from the More… option on the Main menu,
and requires a code before a user can access it.
The Clinician Mode enables the user to:
Verify NIBP calibration and PM dates
Change the alarms to “Permanent Silence” mode
Set target inflation pressure
1.4.1 Accessing the Clinician Mode
This section provides an overview of the Clinician menus. For more
information and details of operation, see the PRO Series 100-400
Monitor Operation Manual.
To enter Clinician Mode from the Main menu, use the rotor control
and LCD.
1. Select the More… button.
2. Select the Service button.