20 DINAMAP® PRO Series 100-400V2 Monitor Pre-Service & Calibration Manual Revision A
2018554-001
1.6.7 Printer Test
This test generates a sample printout from the printer. If no paper is in
the printer, the Monitor generates no alarm.
1. Power on the PRO Monitor and select More...
2. Select the Service button and input 8 3 7 8.
3. From the Super Service Menu, select the Print button,
Ensure the printed test page is clear and easy to read.
1.7 PRO MONITOR CALIBRATION PROCEDURES
Calibration procedures include calibration of the transducers and
adjustment of neonatal and adult overpressure points. These tests
require a manometer and an inflation bulb attached to the PRO
Monitor as shown in Figure 4-1.
The following procedure describes the steps required to calibrate the
pressure transducer. Disassembly is not required.
1.7.1 Calibrating the NIBP Transducer
1. Set up the Monitor and calibration equipment as shown in Fig. 4-1.
2. Power on the PRO Monitor.
3. Verify that calibration equipment reads 0 mmHg of pressure.
Note: If the PRO Monitor displays the language choice menu, select
a language. Select OK and reboot the system.
4. Select the More… button from the Main menu.
5. Select the Service button.
6. Enter the Service Mode access code 2 2 1 3.
7. Select the NIBP button from the Service menu.
Important: From this point, the timing is critical. Before proceeding,
review the following steps.
Setting the Calibration Points
8. Select Cal from the NIBP menu.
9. The Monitor displays “Set Pressure to 0 mmHg.”
10. Ensure calibration test equipment is at 0 mmHg.