16 DINAMAP® PRO Series 100-400V2 Monitor Pre-Service & Calibration Manual Revision A
2018554-001
 
1.6.2 Leak Test
This test performs a leak test of the pneumatic system.
Equipment required:
 CRITIKON
®
 Adult Blood Pressure Cuff (P/N 2774)
 12 foot Gray, Adult/Pediatric Air Hose (P/N 107365)
Figure 4-1
1. Secure the Adult-size cuff to the 12-foot air hose. Attach to the 
PRO Monitor as shown in Figure 4-1. Secure the BP cuff around a 
rigid, unbreakable object that measures at least 2 inches in diameter.
2. Ensure the index-line of the CRITIKON
®
 Blood Pressure Cuff is 
properly aligned within the range-markers on the opposite end of the cuff.
3. Power on the PRO Monitor and select More...
4. Select the Service option and input 8 3 7 8.
5. From the Super Service menu, select the NIBP option.
6. From the NIBP menu, select the LEAK option.
The leak-test sequence closes both valves and turns on the pump. 
The PRO Monitor will self-pressurize the pneumatic test setup to 
approximately 200 mmHg. After settling for 45 seconds, the test 
pressure value will be displayed in mmHg on the LCD. After 30 
seconds, the system displays the final pressure. Confirm that the 
pressure has fallen no more than 5 mmHg.