18 DINAMAP® PRO Series 100-400V2 Monitor Pre-Service & Calibration Manual Revision A
2018554-001
6. Using the inflation bulb, apply the following pressures (measured
by an external digital manometer) and confirm that the Monitor
readings agree with the following table for both PT1 and PT2
channels.
7. If calibration is required, refer to Section 4.7.1.
1.6.4 Neonate Overpressure Check
1. Set up the unit and calibration equipment as shown in Figure 4-1.
2. Power on the PRO Monitor and select More...
3. Select the Service mode and input 2 2 1 3.
4. From the Service menu, select the NIBP button, then select the
Popoff button.
Note: If the overpressure point is out of range, adjust the
overpressure potentiometer as described in 4.7.2.
5. Using the inflation bulb, increase applied pressure until
overpressure occurs. Confirm that pressure at that point is between
150 mmHg to 165 mmHg and system pressure falls to less than 20
mmHg within 4 seconds.
6. Power off the PRO Monitor.
7. If calibration is required, refer to Section 4.7.1.
1.6.5 Adult Overpressure Check
1. Set up the unit and calibration equipment as shown in Figure 4-1.
2. Power on the PRO Monitor and select More...
3. Select the Service mode and input 2 2 1 3.