Revision A DINAMAP® PRO Series 100-400V2 Monitor Pre-Service & Calibration Manual 19
2018554-001
 
4. From the Service menu, select the NIBP button.
5. Select the Check button.
6. Increase applied pressure until overpressure occurs. Confirm that 
pressure at the overpressure is between 300 mmHg and 330 mmHg 
and the system pressure falls to less than 20 mmHg within 8 seconds.
7. Switch off the PRO Monitor and disconnect the calibration kit.
8. If the Monitor fails this test, re-calibrate the unit as described in 
section 4.7.2.
1.6.6 Temperature System Check (PRO 200 & 400)
The PRO Monitor Series 200 & 400 temperature systems use 
ALARIS temperature probes. This system is self-calibrating. The only 
maintenance required is to verify that the temperature functions are 
working properly. These checks require an IVAC probe simulator (P/
N TE 1811), available from ALARIS Medical Systems, Inc., San 
Diego, CA. GE Medical Systems Information Technologies does not 
stock this tester.
To check the temperature system, connect the IVAC probe simulator 
to the temperature probe connector on the front panel, and insert a 
temperature probe into the active holster. 
1. Power on the PRO Monitor.
2. Remove the temperature probe from the probe holster to initiate a 
temperature reading. 
Set the probe simulator to 98.6 and verify that the LCD temperature 
display reads 98.6°F ±1.0°F. The numbers on the temperature LED 
displays should be flashing at this point, indicating the monitor is in a 
real-time monitor mode.
A range of temperatures can be checked, by using the other values 
on the probe simulator (98.0, 80.2, 102.0, and 107.8).
Broken Probe Sensing
Set up the equipment and the probe simulator as in the previous 
procedure. Rotate the temperature selector on the temperature 
simulator to B.P. verify that the Temperature display reads 106.0°F 
±0.2°F. Next. press BROKEN PROBE on the simulator, verify that the 
LED display on the Monitor changes to two dashes, indicating a fault 
condition.
If the PRO Monitor temperature system does not pass these tests, 
the Main PWA needs to be replaced.