10 DINAMAP® PRO Series 100-400V2 Monitor Pre-Service & Calibration Manual Revision A
2018554-001
1.5.3 NIBP Menu
The options on the NIBP menu, pictured above, are used to calibrate
and test BP functions of the PRO Monitor.
Check Button
This button allows the NIBP calibration to be checked. In this mode,
the Monitor functions as a digital manometer. The systolic LED
displays the output of PT1, and the diastolic LED displays the output
of PT2.
With no hose attached, the systolic and diastolic displays will indicate
“000.” See Section 4.7 for a more detailed description of this mode
and the calibration procedure.
Cal Button
This button initiates the calibration procedure. See Section 4.7 for a
step-by-step description of this procedure.
Pop Off Button
This button functions in a similar manner to the Check button, but the
overpressure point is set to the neonate value. See Section 4.7 for a
more detailed description of using this function.
Cycle Button
This button allows the user to activate the Preventative Maintenance
(PM) reminder notice. When you select this button a menu appears,
allowing the user to select the number of days that elapse between
the last NIBP PM date and the activation of the PM reminder screen.
Cycle times range from 30 to 365 days in 5-day increments and OFF.
If you choose OFF as your cycle time, the PM reminder screen is
disabled.