Revision A DINAMAP® PRO Series 100-400V2 Monitor Pre-Service & Calibration Manual 11
2018554-001
During a maintenance check, the current date set on the Monitor is
stored in the eeprom as the last PM date when:
1. An NIBP calibration completes successfully.
2. The Check button under the NIBP button in the Service Mode
Menu is selected.
3. The Pop Off button under the NIBP button in the Service Mode
Menu is selected. If the dates captured by the Check or Pop Off
buttons would ever precede the calibration date, the PM date
would default to the calibration date.
When the number of days since the last PM exceeds the specified
Cycle number of days, a full-screen warning page will be displayed
after the power-up screen, informing the user that preventative
maintenance is due. This screen can be dismissed by pressing any
front panel key, and normal operation will begin.
The date of the last PM and Calibration can be found by selecting the
Info button in the Service Mode Menu or by selecting the Info button
in Clinician Mode Menu.
Preventative Maintenance Reminder Screen
Cancel Button
This button takes the user back to the Service Mode NIBP Menu.
Changes to the Cycle time will not be retained if the Save button has
not been selected before leaving the menu.
Save Button
This button stores the currently displayed Cycle time in the eeprom.
The Cycle time is retrieved for analysis on power-up to determine
whether the Preventative Maintenance Reminder screen should be
shown.
OK Button
This button returns the display to the Main Service menu.