6 DINAMAP® PRO Series 100-400V2 Monitor Pre-Service & Calibration Manual Revision A
2018554-001
Selecting the Silence button displays the dialog box, pictured above,
which mutes all the alarms except the fail-safe alarm. The alarms are
disabled until the Monitor is either powered off and on again or the
Silence button is pressed. Choosing either Yes or No will exit the
menu. If silence is confirmed, the Alarm Silence button lights.
More... Menu Button
The More... menu accesses more options that allow the user to
permanently change the default settings.
Main Button
Select this button to exit the Service menu and return to the Main
menu.
1.4.3 Clinician More... Menu
The Clinician More… menu accesses the options to change the
default settings of several PRO Monitor functions. These options are
summarized below.
Trend Button
Displays a dialog box to automatically clear trend data on power up.
The default setting is Yes. In the default setting, trend data is cleared
when the PRO Monitor is shut off. Select No to retain the trend data
on power-down.
Print Button
Displays a dialog box to restore the print mode on power up to the
user-selected print mode (auto or manual) or the default print mode.
The print mode can be selected through the Print button on the Main
menu. The current mode is displayed in the lower right corner on the
LCD. Select No and the PRO Monitor powers up in manual print
mode. Select Yes, and the monitor retains the user-selected mode.
Set BP Button
Displays a dialog box to power up in a user-selected BP mode (auto/
manual). Select Yes to preserve the user-selected BP mode. Select
No to power up the PRO Monitor in manual mode.
SpO
2
Button
Selecting Yes enters the SpO
2
configuration menu. Selecting No
returns the user to the More… menu. This menu displays a dialog
box that enables the user to adjust the sensitivity of the SpO
2
parameter and to enable the FastSAT Feature (if equipped as a