Revision A DINAMAP® PRO Series 100-400V2 Monitor Pre-Service & Calibration Manual 7
2018554-001
Masimo unit). If the Monitor is equipped for Nellcor SpO
2
, this menu
allows the user to set the Response Mode and SatSeconds feature.
Alarms Button
This button accesses a dialog box to enter the alarms configuration
menu. Selecting Yes enters the menu. Selecting No returns the user
to the More… menu.
Select Reset to return all the alarm limits to the default settings. The
illustration above shows the default settings.
The user can adjust the alarm limits within the ranges specified in the
table below. The high limit must be at least one step higher than the
low limit.
Note: Predictive temperature has no alarm limit alarm.
Select Save to save the displayed settings and return to the More…
menu. Select Cancel to abandon any changes and return to the
More… menu.
Alarm Limits Table
Parameter Range Default
Systolic High 35 - 290 180
Systolic Low 30 - 285 30
Diastolic High 15 - 220 130
Diastolic Low 10 - 215 15
MAP High 25 - 260 140
MAP Low 20 - 255 50
BPM High 35 - 250 150
BPM Low 30 - 245 50
SpO2 High 21 - 100 Off
SpO2 Low 20 - 99 90