Revision A DINAMAP® PRO Series 100-400V2 Monitor Pre-Service & Calibration Manual 15
2018554-001
This is an expected result. To work around this, incrementally step up
or down the settings on your SpO
2
simulator and allow the Monitor to
detect and display the new pulse rate or saturation.
Nellcor recommends use of the SRC-MAX Portable Tester for use
with PRO Monitors equipped with the Nellcor SpO
2
system.
For Monitors equipped with Masimo SpO
2
Masimo recommends BIO-TEK SpO
2
simulators.
1. Connect the appropriate SpO
2
simulator and cable to the SpO
2
connector.
2. Verify the unit displays:
Pulse value
Saturation value
Signal Strength bar graph
3. Disconnect the SpO
2
cable from the simulator.
4. Verify the unit generates an “SpO
2
SENSOR OFF” alarm and
the speaker is sounding.
5. Press the Silence button.
6. Verify the sound has stopped but the error remains displayed.
7. Reconnect the SpO
2
sensor.
8. Verify the unit displays:
Pulse Value
Saturation value
Signal Strength bar Graph