Revision A DINAMAP® PRO Series 100-400V2 Monitor Service Manual A-3
2018554-001
Appendix A
Test Records
Model # __________________ Serial #
__________________________________
Step Description Min Max Actual Pass Fail N/A
5.2 Internal Tests
5.2.2.4
Leak rate
6mmHg / 30sec from 200 mmHg
5.2.2.8 UUT pressure (top) 50 mmHg 46 54
5.2.2.8 UUT pressure (bottom) 50 mmHg 46 54
5.2.2.9 UUT pressure (top) 150 mmHg 145 155
5.2.2.9 UUT pressure (bottom) 150 mmHg 145 155
5.2.2.10 UUT pressure (top) 250 mmHg 244 255
5.2.2.10 UUT pressure (bottom) 250 mmHg 244 255
5.2.2.11 Verify overpressure occurs below 330mmHg
5.3.1 SPO
2
test
5.3.1.3 Verify Bargraph
5.3.1.4 SP0
2
reading at 100 % saturation 96 100
5.3.1.5 SP0
2
reading at 90 % saturation 86 94
5.3.1.6 SP0
2
reading at 80 % saturation 76 84
5.5.1.7 Verify alarm sounds and displays below 90%
5.3.2 BPM test
5.3.2.1 Verify low rate alarm at 50 BPM
5.3.2.2 Heart Rate reading at 50 BPM (SPO
2
) 46 54
5.3.2.4 Heart Rate reading at 120 BPM (SPO
2
) 116 124
5.3.2.8 Heart Rate reading at 80 BPM (SPO
2
) 76 84
5.3.3 NIBP test
5.3.3.4 Initial cuff inflation (Adult) 143 1771
5.3.3.5 Systolic reading (120/80 Adult) 107 133
5.3.3.6 Diastolic reading (120/80 Adult) 67 93
5.3.3.7 Heart rate reading at 80 BPM (NIBP) 76 84
5.3.3.8 Inflate / deflate cycle time <120 sec
5.3.3.12 Initial cuff inflation (Neonate) 94 151
5.3.3.13 Systolic reading (100/65 Neonate) 87 123
5.3.3.14 Diastolic reading (100/65 Neonate) 52 78
5.3.4 Temperature Test
5.3.4.3
Measured temperature in
°
F (98.6 nominal)
98.4 98.8