Innova 2000 Cardiovascular Imaging System
GE Medical Systems
Pre–Installation Manual
REV 5 pim 2337741–100
44
The floor slabs on which the equipment is to be installed must have a
levelness of 1 mm (0.04 in) per meter (40 in) . Baseplate and table basement
must be at the same level.
The preferred installation method for the Innova LC Positioner or the Omega tables is through-bolting.
The through-bolting method can be used in all seismic zones. If through-bolting cannot be used, use
provided floor anchors instead.
Illustration 11 THROUGH–BOLT SUPPLIED (SLAB TYPE FLOOR CONSTRUCTION)
ÉÉÉÉÉÉÉÉÉÉÉÉÉ
ÉÉÉÉÉÉÉÉÉÉÉÉÉ
ÍÍÍ
ÍÍÍ
Cut off end of thru bolt if any interference is
encountered beneath floor
Base plate Through-bolt Flat washer
Steel plate2 hex nuts
Floor
Furnished
through-bolts will
accommodate 200
to 430 mm (8I to
17I) thick floor
FOR EXAMPLE
Floor requirements when using provided floor anchors
The maximum pullout force per provided anchor was calculated assuming:
A concrete compression strength (f’c) of 17.24 MPa at 28 days (which is the minimum required
compression strength).
Anchors installed to the required hole depth of 165.1 mm minimum, and
Center of anchor hole to concrete edge distance 79.4 mm.
Make sure to obtain data on compression strength of the concrete before using floor anchors.
Pan Type Floor Construction Requirement
For Pan type floor construction, steel channels must be designed by a local structural engineer to
span floor joists. See Illustration 12.
Note: For specific floor preparation procedures, refer to im (Installation Manual) 2290880–100
Cardiovascular Imaging System Pre–Installation kit Installation.