Innova 2000 Cardiovascular Imaging System
GE Medical Systems
Pre–Installation Manual
REV 5 pim 2337741–100
45
Illustration 12 – THROUGH–BOLT SUPPLIED (PAN TYPE FLOOR CONSTRUCTION)
ÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉ
Floor
2-hex nuts M20 STL
Flat washer
For pan type floor
construction, joists must
be spanned with steel
channels (customer
furnished)
Furnished thru bolts will accommodate
8 inch (203 mm) to 17 inch (432 mm)
thick floor
1/4 inch (6 mm) to
1/2 inch (13 mm)
grout pad
ÈÈ
ÈÈ
ÈÈ
ÈÈ
ÈÈ
ÈÈ
È
È
È
È
È
È
Spacers
(Customer furnished)
thru bolt M20–500–400Baseplate
FOR EXAMPLE
Illustration 13 LOCATION OF FIXING BOLTS
12 bolts with
a pull out of
3910 N
549.4 2.54
(21.63 0.10) DIA
INNOVA LC
POSITIONER
OMEGA Table
4 bolts with
a pull out of
5650 N
445.3
(17.53)
336
(13.23)
     
TO BOLT HOLES
Isocenter
1242.6 1.5
(48.92 0.06)
COMPACT
& LONG
1125.7 1.5
(44.32 0.06)
LONG
(optional)
For alternative table bolts or seismic area, refer to template drawing shown in Illustration 43.
Hole dimension and preferred location in concrete floor
If in the examination room, the Innova LC Positioner and the OMEGA patient tables
are not placed on a computer floor but directly put on concrete floor, the location of
the cable access needs to be carefully planned.
Otherwise, if the cable run is located under the concrete floor, the cables will have to
come through the floor and in this case you will need two holes, one for the LC
Positioner and the other for the patient table.
The diameter of both holes can be the same (225 mm / 9 in).