General Procedures and Functional Checks
4-38 LOGIQ V3/V5/V5 Expert – Basic Service Manual
5726264-100 English Rev.8
4-3-26 Basic Measurements
NOTE: The following instructions assume that you first scan the patient
and then press Freeze.
4-3-26-1 Check Distance and Tissue Depth Measurement
1. Press Measure once to display an active caliper.
2. Move the trackball to position the active caliper at the start
point (distance) or the most anterior point (tissue depth).
3. Press Set to fix the start point.
4. The system fixed the first caliper and displays a second
active caliper.
5. Move the trackball to position the second active caliper at
the end point (distance) or the most posterior point (tissue
depth).
6. Press Set to complete the measurement. The system
displays the distance or tissue depth value in the
measurement results window.
Before you complete a measurement:
To toggle between active calipers, rotate Cursor Select button.
To erase the second caliper and the current data measured and
start the measurement again, press Clear.
NOTE: To rotate through and activate previously fixed calipers, rotate
Cursor Select button.
NOTE: After you complete the measurement, to erase all data that has
been measured to this point, but not data entered onto
worksheets, press Clear.