EasyManuals Logo

GE LOGIQ V3 User Manual

GE LOGIQ V3
305 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #278 background imageLoading...
Page #278 background image
Care and Maintenance
10-14 LOGIQ V3/V5/V5 Expert – Basic Service Manual
5726264-100 English Rev.8
10-6-2-2 Peripheral/option checks
If any peripherals or options are not part of the system
configuration, the check can be omitted.
Refer to the User Manual for a list of approved peripherals/
options.
10-6-2-3 Mains cable inspection
Table 10-7: GE approved peripheral/hardware option functional checks
Step Item Description
1. Media Verify media drive(s) read/write properly. Clean if necessary.
2. B/W Printer Verify hardcopy output of the B/W video page printer. Clean heads and
covers if necessary.
3. Color Printer Verify hardcopy output of the Color video page printer. Clean heads and
covers if necessary.
4. DICOM Verify that DICOM is functioning properly. Send an image to a DICOM
device.
5. ECG Verify basic operation with customer
6. Footswitch Verify that the footswitch is functioning as programed. Clean as necessary.
7. DVD Verify that the DVD is functioning properly. Clean heads and covers if
necessary.
Table 10-8: Mains Cable Inspection, As Appropriate
Step Item Description
1. Unplug Cord Disconnect the mains cable from the wall and Ultrasound system.
2. Inspect Inspect it and its connectors for damage of any kinds.
3. Verify Verify that the LINE, NEUTRAL and GROUND wires are properly attached
to the terminals, and that no strands may cause a short circuit.
4. Verify Inlet connector retainer is functional.

Table of Contents

Other manuals for GE LOGIQ V3

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ V3 and is the answer not in the manual?

GE LOGIQ V3 Specifications

General IconGeneral
BrandGE
ModelLOGIQ V3
CategoryMedical Equipment
LanguageEnglish

Related product manuals