Manual Defibrillation / Non-Synchronized Defibrillation
227 490 02-C Marquette Responder® 3000 23
Warning
Risk of Skin Burns / Equipment Damage — Do not
attach the pads over
-
sternum or clavicle
-
nipples
-
implanted pacemaker or defibrillator devices.
3
2
Analyse
Sync
Figure 4-8. Buttons to initiate defibrillator
charging and to trigger the shock
correct
wrong
Figure 4-9. Removing the pads
Before delivering the shock, check that the pads
are firmly seated.
Defibrillate the patient as described for
defibrillation with paddles (page 19).
Be sure to charge the defibrillator with
and
to deliver the shock by simultaneously pressing
the buttons
and
(Figure 4-8).
Carefully remove the electrodes after use
(Figure 4-9) and discard them immediately.
Note
Discard disposable pads immediately after use. Do
not reuse them.