Manual Defibrillation / Non-synchronized Defibrillation
24 Marquette Responder® 3000 227 490 02-C
Note
If you are using internal electrodes with the
Marquette Responder® 3000, defibrillator charg-
ing must be initiated with button and the
shock is delivered by simultaneously pressing
buttons and .
Figure 4-10. External counter electrode
Figure 4-11. Inserting the spoon electrode
Using Internal Defibrillation Elec-
trodes
Warning
Shock Hazard – Always switch off the device
before exchanging the defibrillation electrodes
and internal spoons.
Spoon-shaped electrodes are used for internal
defibrillation. Their contact surface must match
the dimensions of the heart. The spoons must make
full contact with the heart. There is a choice of 3
different spoon sizes. You can use either two
spoon electrodes or one spoon electrode and one
external counter electrode for defibrillation
(Figure 4-10, chapter 20 "Order Information").
Sterilize internal electrodes before each use
(chapter 18 "Cleaning, Maintenance").
Inserting the Spoon Electrode
Screw the counter nut
2
(Figure 4-11) onto the
electrode as far as it will go.
Screw the contact paddle
1
into the handle as
far as it will go, then bring it into the appropri-
ate position.
Now fix the contact paddle by screwing the
counter nut
2
tight against the handle
3
.