Manual Defibrillation / Non-Synchronized Defibrillation
227 490 02-C Marquette Responder® 3000 25
3
2
Analyse
Sync
Figure 4-12. Buttons to initiate defibrillator
charging and to trigger the shock
Defibrillating the Patient
With internal electrodes, it is not possible to select
a value above 50 Joules, because higher energies
may damage the myocardium. When you set the
energy selector to a value above 50 Joules, you
will be alerted by a message and defibrillator
charging will not proceed.
Defibrillate the patient as described for
defibrillation with paddles (page 19).
Be sure to charge the defibrillator with
and
to deliver the shock by simultaneously pressing
buttons
and (Figure 4-12).