Loading...
Semiautomatic Defibrillation
32 Marquette Responder® 3000 227 490 02-C
ECG
Pads
15.07.1999 09:05:00
160 / 40
bpm
semiautom.
200 J
161
ECG
SpO2
etCO2
QRSPulse
Tone
OFF
Next
Menu
e
a c db
Figure 5-1. Standard display
a Heart rate
green = HR detection OK
blue = lead failure
red = medical alarm
b QRS blip
c semiautomatic mode selected
d selected energy
e menu
Defibrillating the Patient in the
Semiautomatic Mode
We recommend to defibrillate the patient with
disposable adhesive pads in this mode. Thus, you
only have to apply two electrodes, because the
ECG signal can also be acquired via the pads.
If you are using separate ECG electrodes, please
refer to the end of this chapter.
Apply the defibrillation pads as described on
page 22.
Before delivering the shock, check that the pads
are firmly seated.
Set the energy selector to "Autoseq".
Check that the semiautomatic mode is enabled
(
c
, Figure 5-1) and that the selected energy is
displayed (
d
, Figure 5-1).