Semiautomatic Defibrillation
34 Marquette Responder® 3000 227 490 02-C
Note
-
Immediately after selecting "Autoseq", the
charge sequence restarts with the lowest en-
ergy level. There is no time limit for the Auto-
sequence shocks. The Autosequence can only
be interrupted by activation of the energy se-
lector.
The energy level selected for the 3rd shock is
maintained for all subsequent shocks.
-
If the shock is not delivered within 30 seconds
of charging, an automatic internal safety dis-
charge will be initiated. In this case you must
press the
Analyse
button again.
Note
If the analysis algorithm does not recommend a
defibrillation, even though the patient is suspected
to suffer from a shockable arrhythmia, press
Analyse
again. If possible, apply ECG electrodes to acquire
the ECG.
Also the device can be switched to manual defi-
brillation by authorized personnel.
Shock
(198 J)
17.07.1999 09:30:26 HR:161 bpm
Patient………………..…. :
Date of birth…...…….…. :
User.......……………..…. :
Comments....……….….. :
Defib Mode……………...:
Selected Energy…..…....:
Delivered Energy............ :
No of Shocks………..…. :
HR Alarm Limits... ……...:
ETC Alarm Limits…...…. :
HELLIGE
semiautom./button
200 J
198 J
1
40 :160 bpm
20 : --- mmHg
25 mm/s 1 cm/mV 50Hz 35Hz
Pads
a b c d
eg
f
Figure 5-5. Example of a recording initiated by a defibrillation shock
a
reason for recording, delivered energy
b
date, time
c
heart rate
d
intervention report
e
user text
f
speed, gain, filter(s)
g
shock delivery, 5 seconds blanked