Defibrillator Setup
66 Marquette Responder® 3000 227 490 02-C
Event Texts
Entry of up to 8 event texts (max. 8 characters
each); these texts are assigned to the softkeys F1
to F4 (chapter 3, section "Event Button")
Using
←
or
→
, choose the position for the
character; with
↑
or
↓
, choose the letter or
number.
Battery
For details on the battery maintenance program,
please refer to chapter 14, section "Battery
Maintenance Program".
Options
For activation of optional program modules
(Figure 13-6).
The defibrillator serial number has already been
entered. We need to know this number when you
order optional program modules. Then you will be
given the option code number. When you enter this
number and exit the configuration menu with
Save & EXIT
, the defibrillator will automatically
restart and activate the new program module(s).
ECG
I
20.01.1999 14:35:37
160 / 40
bpm
Previous
Menu
–+
91
OPTIONS
SER. NO
12SL
VFVT
100.00.309
604.705.273.662
221.763.542.440
Figure 13-6. Activating optional program
modules
12SL: ECG analysis program
VFVT: semiautomatic defibrillator