EasyManua.ls Logo

GE Masoneilan 12420 - 9.3. Removal of 12400 Case and Torque Tube Sub-Assembly

GE Masoneilan 12420
80 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
50 | GE Oil & Gas ® 2014 General Electric Company. All rights reserved.
9.3. REMOVAL OF 12400 CASE AND TORQUE TUBE SUB-ASSEMBLY (SEE FIGURES 22 & 23)
a. Switch off power supply.
b. For instruments with a displacer chamber, close isolation valves, and purge the chamber.
c. Remove the upper flange (146) and blind flange (144).
d. Lower the torque arm (135) and unhook the displacer (130). A hook-shaped 3 mm steel wire facilitates unhooking and
holding of the displacer. The wire may be inserted through the clevis hole.
e. Remove the two torque arm screws (133) and remove the torque arm (135) from the chamber.
f. Remove the displacer from its chamber (131) or tank.
g. Be sure that requirements for instruments installed in explosive areas are strictly followed. Remove screw (106) from
connection compartment and unscrew cover (104). Disconnect supply electrical wires and other equipment from
terminals (90).
h. Remove nuts (142) holding the torque tube subassembly and slide the subassembly out of the mechanism chamber.
PART REFERENCE
130
Displacer
143
Gasket
131
Displacer chamber
144
Blind flange
132 Knife of torque tube 145 Stud
133
Torque arm screw
146
Upper flange
134
Torque arm screw block
147
Upper flange stud
135
Torque arm
148
Upper flange stud nut
137
Torque tube housing S/A
149
Upper flange gasket
142 Stud nut
CAUTION: To remove the displacer, the torque arm must be removed. When the two arm srews (133) are loosened,
hold the torque arm so as not to damage the torque tube (Figure 22).
Figure 22
Figure 23

Table of Contents

Related product manuals