MA-014 - TRANSFIX DGA 500 User Guide Rev 1.1 13-Oct-15 Page 19 of 28
Code
Error
Description
27
Temperature Sensor(s) 1 disconnected
The product is not receiving a signal from one or more of
the vital temperature sensors. The product will abort the
measurement.
28
Temperature Sensor(s) 2 disconnected
The product is not receiving a signal from one or more of
the less vital temperature sensors. The product will
complete the measurement. The final values may not be
reliable.